FDA Adverse Event Malfunction Summary report: N

SYNCHROMED II

MDR report key: 1940669 · Received December 17, 2010

Report

Report Number
3004209178-2010-10554
Event Type
Malfunction
Date Received
December 17, 2010
Date of Event
November 29, 2010
Report Date
November 29, 2010
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LKK
PMA / PMN Number
P860004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE SPINAL SEGMENTS OF 3 CATHETERS HAVE BEEN RETURNED TO THE MANUFACTURER FOR ANALYSIS. A FOLLOW-UP REPORT WILL BE SENT WHEN THE ANALYSIS IS COMPLETE.

Description of Event or Problem · 1

DURING IMPLANT, THE PHYSICIAN HAD DIFFICULTY PLACING THE CATHETERS. INITIALLY DURING THE PROCEDURE, THE PHYSICIAN GAINED ACCESS TO THE INTRATHECAL SPACE UTILIZING A VERY STEEP APPROACH ANGLE (APPROXIMATELY 85 DEGREES FROM THE SPINE) AND ATTEMPTED TO PLACE THE SPINAL SEGMENT OF (B)(4). THE CATHETER APPEARED TO LOOP ON THE FLUOROSCOPIC IMAGE AND STARTED TO TRAVEL IN THE CAUDAL DIRECTION. THE PHYSICIAN THEN REMOVED THE NEEDLE AND CATHETER AND ATTEMPTED A SLIGHTLY SHALLOWER ANGLE IN THE SAME INTERLAMINAR SPACE (L3-4) THIS TIME WITH (B)(4) WITH THE SAME RESULT. THIS CATHETER WAS REMOVED AND THE SAME WAS ATTEMPTED USING THE (B)(4) AND AGAIN THE SAME RESULT WAS SEEN. A NEW CATHETER WAS ATTEMPTED EACH TIME AS EACH HAD BECOME SOMEWHAT DEFORMED BY THE ATTEMPTS TO PLACE. FINALLY, THE NEEDLE WAS REMOVED AND REPLACED AT A SHALLOW PARAMEDIAN APPROACH IN THE L2-L3 SPACE. THIS TIME, THE SPINAL SEGMENT OF (B)(4) WAS PLACED WITHOUT COMPLICATION. THE 7.6 CM PUMP SEGMENT WAS DISCARDED. THE IMPLANTED PORTION OF THE (B)(4) (32.2 CM) WAS ATTACHED TO THE 66 CM PUMP SEGMENT OF (B)(4). THERE WAS REPORTED TO BE NO INJURY. THE PATIENT RECOVERED WITHOUT SEQUELA. THE DEVICE SYSTEM WAS GOING TO BE USED TO DELIVER LIORESAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHROMED II LKK MDT PUERTO RICO OPERATIONS CO., JUNCOS 8637-40 NA

Patients

Seq Age Sex Outcome Treatment
1 50 YR CATHETER: MODEL 8709, LOT# N250773004| CATHETER: MODEL 8731SC, LOT# N274062005| CATHETER: MODEL 8731SC, LOT# N274062004| IMPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| IMPLANTED:| IMPLANTED: