VASOVIEW 7 XB EVH SYSTEM
Report
- Report Number
- 2242352-2010-03787
- Event Type
- Injury
- Date Received
- December 22, 2010
- Date of Event
- December 2, 2010
- Report Date
- December 3, 2010
- Manufacturer
- MAQUET CARDIOVASCULAR LLC
- Product Code
- GEI
- PMA / PMN Number
- K041981
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION SUMMARY: A VISUAL INSPECTION REVEALED THAT THE DISSECTION TIP WAS RECEIVED WITH THE CONICAL TIP ATTACHED, BUT IT COULD BE REMOVED WITH MINIMAL FORCE. THE TIP FORMED A COMPLETE ASSEMBLY. THERE WAS SOME ADHESIVE VISIBLE ON THE CONNECTING SURFACES. THERE WAS SOME EVIDENCE OF BLOOD IN THE TIP. BASED UPON THIS OBSERVATION, THE REPORTED COMPLAINT "DISSECTION TIP SEPARATED" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).
THE HOSP REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE DISSECTION TIP SEPARATED FROM THE MAIN PART IN THE PT. IT WAS RETRIEVED THROUGH THE ORIGINAL INCISION. A NEW KIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASOVIEW 7 XB EVH SYSTEM | ENDOSCOPIC VESSEL HARVESTING SYSTEM | GEI | MAQUET CARDIOVASCULAR LLC | VH-3200 | 25020285 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |