LAMITRODE 88 60CM SURGICAL LEAD
Report
- Report Number
- 1627487-2010-03873
- Event Type
- Injury
- Date Received
- December 21, 2010
- Date of Event
- November 22, 2010
- Report Date
- November 22, 2010
- Manufacturer
- ST JUDE MEDICAL - NEUROMODULATION DIV
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: REVIEW OF THE DEVICE HISTORY RECORD FOUND A NONCONFORMANCE; HOWEVER, THE NONCONFORMANCE WAS IDENTIFIED AS A COSMETIC ISSUE AND DID NOT AFFECT PRODUCT INTEGRITY OR PRODUCT FUNCTIONALITY AND WAS APPROVED FOR USE. THE DHR ANOMALY IS NOT RELATED TO THE ALLEGED DEVICE FAILURE. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REFERENCE MFR. REPORT# 1627487-2010-03872.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAMITRODE 88 60CM SURGICAL LEAD | SPINAL CORD STIMULATION LEAD | LGW | ST JUDE MEDICAL - NEUROMODULATION DIV | 3288 | A983011993 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |