FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY (WITH T:CONNECT MOBILE APP)

MDR report key: 19404601 · Received May 28, 2024

Report

Report Number
3013756811-2024-94406
Event Type
Malfunction
Date Received
May 28, 2024
Date of Event
May 6, 2024
Report Date
May 28, 2024
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00389152000107
PMA / PMN Number
K203234
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PER THE TANDEM USER GUIDE: THE T:SLIM X2 WITH CIQ TECHNOLOGY USER GUIDE STATES: DO NOT USE ANY OTHER INSULIN WITH YOUR SYSTEM OTHER THAN U-100 HUMALOG OR U-100 NOVOLOG. ONLY HUMALOG AND NOVOLOG HAVE BEEN TESTED AND FOUND TO BE COMPATIBLE FOR USE IN THE SYSTEM. USE OF INSULIN WITH LESSER OR GREATER CONCENTRATION CAN RESULT IN UNDER DELIVERY OR OVER DELIVERY OF INSULIN. THIS CAN CAUSE VERY HIGH OR A VERY LOW BLOOD GLUCOSE. THE PUMP USER GUIDE CAUTIONS THAT INSULIN SHOULD BE AT ROOM TEMPERATURE BEFORE USE OR AIR BUBBLES COULD FORM IN THE CARTRIDGE. THE PUMP USER GUIDE INSTRUCTS THE USER TO REMOVE ANY RESIDUAL AIR FROM THE CARTRIDGE. H3 OTHER TEXT : DEVICE NOT RETURNED.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN OCCLUSION ALARM OCCURRED. REPORTEDLY, THE CUSTOMER WAS USING COLD FIASP INSULIN WITH THE PUMP, AND WAS NOT PERFORMING THE AIR REMOVAL STEP OF THE LOAD PROCESS PROPERLY. TANDEM CUSTOMER TECHNICAL SUPPORT INFORMED CUSTOMER OF PROPER INSULIN/SUPPLY USAGE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
664030 T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY (WITH T:CONNECT MOBILE APP) ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1000354 00389152000107

Patients

Seq Age Sex Outcome Treatment
1 55 YR Male INSULIN: FIASP.