FDA Adverse Event Malfunction Summary report: N

INFINITI VISION SYSTEM OZIL

MDR report key: 1940431 · Received December 17, 2010

Report

Report Number
2028159-2010-02400
Event Type
Malfunction
Date Received
December 17, 2010
Report Date
November 17, 2010
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K082845
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

A SAMPLE HAS BEEN RETURNED TO THE MFR, BUT THE ANALYSIS HAS NOT YET BEEN COMPLETED. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADD'L REPORTABLE INFO BECOMES AVAILABLE. (B)(4).

Description of Event or Problem · 1

A NURSE REPORTED THAT THERE WAS NO VACUUM IN VITRECTOMY MODE. THE SYSTEM WAS EXCHANGED AND THE VITRECTOMY WAS COMPLETED WITHOUT DELAY IN THE CASE. ADD'L INFO HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITI VISION SYSTEM OZIL PHACOFRAGMENTATION SYSTEM HQC ALCON - IRVINE TECHNOLOGY CENTER INFINITI OZIL NA

Patients

Seq Age Sex Outcome Treatment
1 NI