FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 1940412 · Received December 9, 2010

Report

Report Number
1824206-2010-11490
Event Type
Malfunction
Date Received
December 9, 2010
Date of Event
November 15, 2010
Report Date
November 15, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH REPLACED THE HYDRAULIC VALVE SOLENOID TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE HEAD SECTION WILL NOT RISE DUE TO A CONTAMINATED VALVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1