TUBE, GASTROINTESTINAL (AND ACCESS)
Report
- Report Number
- 3006260740-2010-00378
- Event Type
- Injury
- Date Received
- December 20, 2010
- Date of Event
- November 12, 2010
- Report Date
- November 29, 2010
- Manufacturer
- C. R. BARD INC. (BASD)
- Product Code
- KGC
- PMA / PMN Number
- K904779
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE COMPLAINT OF A BREAK IN THE RETENTION DOME IS CONFIRMED, USER RELATED. UPON RECEIPT A PARTIAL SEMI-CIRCULAR BREAK IN THE RETENTION DOME WAS OBSERVED. A SMALL SECTION OF THE BROKEN DOME IS STILL ATTACHED TO THE FEEDING TUBE AND REVEALS A COMPLETE BOND. THE DULL AND ROUNDED VENEER IDENTIFIED AT THE DOME SITE ARE TRAITS THAT ARE INDICATIVE OF EXCESSIVE FORCE THAT WAS APPLIED DURING THE REMOVAL OF THIS DEVICE. THE LACK OF ANY ASSOCIATED DAMAGE SUGGESTS THE TEAR IN THE DOME WAS CAUSED BY STRETCHING THE DOME MATERIAL BEYOND ITS ELASTIC LIMITS DURING REMOVAL. GROSS VISUAL AND MICROSCOPIC EXAMINATIONS SHOW NO EVIDENCE OF A DEFECT OR DEFORMITY RELATED TO THE MANUFACTURING PROCESS. A CHR IS NOT POSSIBLE, AS NO MANUFACTURING LOT NUMBER INFO HAS BEEN PROVIDED BY THE COMPLAINANT.
A BREAK IN THE RETENTION DOME DURING TRACTION REMOVAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TUBE, GASTROINTESTINAL (AND ACCESS) | KGC | C. R. BARD INC. (BASD) | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |