FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 19402677 · Received May 28, 2024

Report

Report Number
9610877-2024-53990
Event Type
Malfunction
Date Received
May 28, 2024
Date of Event
May 10, 2024
Report Date
May 28, 2024
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
UDI-DI
04961333070909
PMA / PMN Number
K951196
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR.OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CFB (COHERENT FIBER BUNDLE) BROKEN.BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CFB (COHERENT FIBER BUNDLE).IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE INSERTION FLEXIBLE TUBE CRUSHED, AND THE LIGHT GUIDE CABLE CRUSHED; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT.BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
676585 PENTAX FIBER NASO PHARYNGO LARYNGOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE FNL-7RP3 04961333070909

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown