FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 19402677
·
Received May 28, 2024
Report
- Report Number
- 9610877-2024-53990
- Event Type
- Malfunction
- Date Received
- May 28, 2024
- Date of Event
- May 10, 2024
- Report Date
- May 28, 2024
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- EOB
- UDI-DI
- 04961333070909
- PMA / PMN Number
- K951196
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR.OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CFB (COHERENT FIBER BUNDLE) BROKEN.BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CFB (COHERENT FIBER BUNDLE).IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE INSERTION FLEXIBLE TUBE CRUSHED, AND THE LIGHT GUIDE CABLE CRUSHED; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT.BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 676585 | PENTAX | FIBER NASO PHARYNGO LARYNGOSCOPE | EOB | HOYA CORPORATION PENTAX TOKYO OFFICE | FNL-7RP3 | 04961333070909 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |