FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 19402660 · Received May 28, 2024

Report

Report Number
9610877-2024-53979
Event Type
Malfunction
Date Received
May 28, 2024
Date of Event
May 7, 2024
Report Date
May 28, 2024
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL VNL-1570STK IS AVAILABLE IN THE USA WITH A 510K NUMBER K162151. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR.OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE DISTAL END WITH CCD MODULE BLACKOUT.BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE DISTAL END WITH CCD MODULE.BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT ).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
657313 PENTAX VIDEO NASO PHARYNGO LARYNGOSTROBOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE VNL-1070STK

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown