FDA Adverse Event Malfunction Summary report: N

TERUMO ADVANCED PERFUSION SYSTEM 1

MDR report key: 1940127 · Received December 23, 2010

Report

Report Number
1828100-2010-02740
Event Type
Malfunction
Date Received
December 23, 2010
Date of Event
December 8, 2010
Report Date
December 23, 2010
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DTQ
PMA / PMN Number
K022947
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

DURING USE OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE USER REPORTED AN INTERMITTENT ALARM ON THE LEVEL SENSOR. THE USER REPORTED THE LEVEL MODULE STAYED GREEN DURING THE ALARM, AND THERE WAS NO ALARM INDICATION ON THE SCREEN. THE LEVEL SYSTEM WAS TURNED OFF, AND THE DEVICE WAS USED TO CONCLUDE THE PROCEDURE. THE USER REPORTED THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT DURING THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TERUMO ADVANCED PERFUSION SYSTEM 1 HEART LUNG CONSOLE DTQ TERUMO CARDIOVASCULAR SYSTEMS CORP. 801763

Patients

Seq Age Sex Outcome Treatment
1