MONO BENDABLE ONE PIECE IMPLANT
Report
- Report Number
- 3011390931-2024-00236
- Event Type
- Injury
- Date Received
- May 27, 2024
- Date of Event
- November 7, 2023
- Report Date
- May 27, 2024
- Manufacturer
- NORIS MEDICAL LTD
- Product Code
- DZE
- UDI-DI
- 07290108818345
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
AN INVESTIGATION HAS BEEN COMPLETED, EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF MISSING EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC. ) / PATIENT HABITS (E.G., SMOKING), AND REASON FOR IMPLANT REMOVAL. THE DHR WAS REVIEWED FOR LOT#7000994, AND NO EVIDENCE WAS DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. TO DATE, ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
UDI RELATED DATA QUALITY UPDATES ONLY.
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022. ON (B)(6) 2023, LOOSENING OF IMPLANT WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1237790 | MONO BENDABLE ONE PIECE IMPLANT | MONO BENDABLE ONE PIECE DENTAL IMPLANT | DZE | NORIS MEDICAL LTD | NMBV3310 | 7000994 | 07290108818345 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Female |