FDA Adverse Event Malfunction Summary report: N

ADULT DUAL-HEATED BREATHING CIRCUIT

MDR report key: 1940002 · Received December 28, 2010

Report

Report Number
9611451-2010-00795
Event Type
Malfunction
Date Received
December 28, 2010
Date of Event
November 18, 2010
Report Date
December 1, 2010
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RT204 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: THE LATCH STRENGTH OF THE ELBOW CONNECTOR ON THE RETURNED INSPIRATORY LIMB OF THE COMPLAINT BREATHING CIRCUIT WAS TESTED USING A FORCE GAUGE. RESULTS: THE LATCH STRENGTH ON THE CONNECTOR WAS FOUND TO BE LESS THAN ON A NORMAL BREATHING CIRCUIT, CONFIRMING THE REPORTED FAULT THAT THE CONNECTOR WAS DAMAGED. A LOT CHECK COULD NOT BE COMPLETED AS NO LOT NUMBER WAS PROVIDED. CONCLUSION: OUR RESULTS SHOW THAT THE LATCH STRENGTH BETWEEN THE HEATER WIRE PLUG AND THE ELBOW CONNECTOR WAS OUTSIDE PRODUCT SPECIFICATIONS, WHICH LIKELY RESULTED IN THE PARTS BECOMING LOOSE AS REPORTED. WE WERE UNABLE TO DETERMINE THE CAUSE OF THE LOOSE CONNECTOR. A LEAK IN THE SYSTEM WILL USUALLY BE DETECTED BY AN ALARM ISSUING FROM THE VENTILATOR. THE RT204 USER INSTRUCTIONS STATE THE FOLLOWING WARNINGS: CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE; PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; SET APPROPRIATE VENTILATOR ALARMS. FISHER & PAYKEL HEALTHCARE HAS RECEIVED ONE OTHER COMPLAINT OF THIS NATURE IN THE PAST YEAR TO THE END OF NOVEMBER 2010.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT A CONNECTOR ON AN RT204 BREATHING CIRCUIT CRACKED PRIOR TO USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADULT DUAL-HEATED BREATHING CIRCUIT BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED RT204 NOT PROVIDED

Patients

Seq Age Sex Outcome Treatment
1