FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED BREATHING CIRCUIT

MDR report key: 1939957 · Received December 28, 2010

Report

Report Number
9611451-2010-00791
Event Type
Malfunction
Date Received
December 28, 2010
Report Date
December 1, 2010
Manufacturer
FISHER & PAYKEL HEALTHCARE
Product Code
BTT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE BREATHING CIRCUIT WAS CHECKED FOR BENT OR BROKEN HEATER WIRE PINS. RESULTS: ONE OF THE INSPIRATORY HEATER WIRE PINS WAS SPLIT, PREVENTING FULL INSERTION OF A HEATER WIRE ADAPTOR. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 100412. CONCLUSION: ALL BREATHING CIRCUITS ARE TESTED FOR CONNECTIVITY AND ELECTRICAL CONTINUITY DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. IT IS POSSIBLE FOR THE USER TO DAMAGE THE HEATER WIRE PINS DURING USE IF THE HEATER WIRE ADAPTOR IS INSERTED INTO THE HEATER WIRE PLUG ON AN ANGLE. FOR DAMAGED PINS REPORTED TO US BY HEALTHCARE FACILITIES, IT IS IMPOSSIBLE FOR US TO DETERMINE WHETHER THE PINS WERE DAMAGED DURING PRODUCTION OR BY THE END USER. DAMAGED PINS DO NOT PRECLUDE VENTILATION OF THE PATIENT, BUT PREVENT THE HEATING OF THE GAS DELIVERED. (B)(4).

Description of Event or Problem · 1

A DISTRIBUTOR IN (B)(6) REPORTED THAT AN RT200 ADULT BREATHING CIRCUIT HAD A BROKEN HEATER WIRE PIN. THE BROKEN PIN WAS FOUND BY THE DISTRIBUTOR PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADULT DUAL HEATED BREATHING CIRCUIT BTT BTT FISHER & PAYKEL HEALTHCARE RT200 100412

Patients

Seq Age Sex Outcome Treatment
1