ADULT DUAL HEATED BREATHING CIRCUIT
Report
- Report Number
- 9611451-2010-00791
- Event Type
- Malfunction
- Date Received
- December 28, 2010
- Report Date
- December 1, 2010
- Manufacturer
- FISHER & PAYKEL HEALTHCARE
- Product Code
- BTT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). METHOD: THE BREATHING CIRCUIT WAS CHECKED FOR BENT OR BROKEN HEATER WIRE PINS. RESULTS: ONE OF THE INSPIRATORY HEATER WIRE PINS WAS SPLIT, PREVENTING FULL INSERTION OF A HEATER WIRE ADAPTOR. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 100412. CONCLUSION: ALL BREATHING CIRCUITS ARE TESTED FOR CONNECTIVITY AND ELECTRICAL CONTINUITY DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. IT IS POSSIBLE FOR THE USER TO DAMAGE THE HEATER WIRE PINS DURING USE IF THE HEATER WIRE ADAPTOR IS INSERTED INTO THE HEATER WIRE PLUG ON AN ANGLE. FOR DAMAGED PINS REPORTED TO US BY HEALTHCARE FACILITIES, IT IS IMPOSSIBLE FOR US TO DETERMINE WHETHER THE PINS WERE DAMAGED DURING PRODUCTION OR BY THE END USER. DAMAGED PINS DO NOT PRECLUDE VENTILATION OF THE PATIENT, BUT PREVENT THE HEATING OF THE GAS DELIVERED. (B)(4).
A DISTRIBUTOR IN (B)(6) REPORTED THAT AN RT200 ADULT BREATHING CIRCUIT HAD A BROKEN HEATER WIRE PIN. THE BROKEN PIN WAS FOUND BY THE DISTRIBUTOR PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT DUAL HEATED BREATHING CIRCUIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE | RT200 | 100412 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |