CD HORIZON SOLERA 5.5/6.0
Report
- Report Number
- 1030489-2024-00569
- Event Type
- Malfunction
- Date Received
- May 27, 2024
- Date of Event
- February 26, 2024
- Report Date
- May 27, 2024
- Manufacturer
- MEDTRONIC SOFAMOR DANEK USA, INC
- Product Code
- HXX
- UDI-DI
- 00613994429742
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION SUMMARY: PRODUCT ANALYSIS #804314455: PART # 5480004V LOT # K22L3316 VISUAL EXAMINATION CONFIRMED THE INSTRUMENT TIP HAS BROKEN OFF AND NOT RETURNED FOR ANALYSIS. OPTICAL INSPECTION OF THE FRACTURE SURFACE REVEALED A FAIRLY FLAT FRACTURE SURFACE WITH CIRCULAR MATERIAL FLOW. THIS TYPE OF DAMAGE IS CONSISTENT WITH TORSIONAL OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING THE DEVICES USED FOR SPINAL THERAPY. IT WAS REPORTED THAT THE INSTRUMENTS BROKE DURING NORMAL INSTRUMENT SERVICING. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 329614 | CD HORIZON SOLERA 5.5/6.0 | SCREWDRIVER | HXX | MEDTRONIC SOFAMOR DANEK USA, INC | 5480004V | K22L3316 | 00613994429742 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |