FDA Adverse Event Malfunction Summary report: N

FREESTYLE FREEDOM

MDR report key: 1939869 · Received December 28, 2010

Report

Report Number
2954323-2010-01874
Event Type
Malfunction
Date Received
December 28, 2010
Date of Event
December 6, 2010
Report Date
April 22, 2011
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. NOTE: THE METER IS DESIGNED TO DISPLAY READINGS OF 20 MG/DL TO 500 MG/DL. THIS METER DOES NOT SHOW A NUMERIC VALUE GREATER THAN 500 MG/DL. A BLOOD GLUCOSE READING ABOVE 500 MG/DL WILL READ AS A "HI" IN THE ADC METER.

Additional Manufacturer Narrative · 1

AN INVESTIGATION WAS PERFORMED ON RETURNED METER (B)(4) AND RETURNED TEST STRIPS LOT NUMBER 1022117. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL FUNCTIONAL TEST RESULTS WERE WITHIN RANGE SPECIFICATION. NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THIS IS A FINAL REPORT.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER WITHIN 10 MINUTES. RESULTS OF 501 MG/DL AND 127 MG/DL WERE PLOTTED ON THE PARKES ERROR GRID. THE RESULTS FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.