FDA Adverse Event Malfunction Summary report: N

ENDOWRIST

MDR report key: 19397889 · Received May 24, 2024

Report

Report Number
2955842-2024-14729
Event Type
Malfunction
Date Received
May 24, 2024
Date of Event
April 12, 2024
Report Date
April 30, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874112380
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE LARGE CLIP APPLIER INSTRUMENT AND COMPLETED THE DEVICE EVALUATION. THE INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A BROKEN GRIP CABLE AT THE IDLER PULLEYS. THE CABLE WAS FULLY BROKEN. THERE WAS NO EVIDENCE OF DISCOLORATION OR CORROSION/CONTAMINATION ON THE CABLES TO INDICATE A REPROCESSING INDUCED ISSUE. THE COMPONENTS SURROUNDING THE BROKEN CABLE DID NOT EXHIBIT ABNORMAL SHARP EDGES OR DAMAGE THAT WOULD HAVE CONTRIBUTED TO THE CABLE BREAKING. A REVIEW OF THE PROVIDED IMAGE WAS PERFORMED BY AN ISI FAILURE ANALYSIS ENGINEER (FAE). THE FOLLOWING ADDITIONAL INFORMATION WAS PROVIDED: IT LOOKED LIKE A BROKEN GRIP CABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED CHOLECYSTECTOMY SURGICAL PROCEDURE, THE LARGE CLIP APPLIER INSTRUMENT WIRE BROKE WHILE THE CUSTOMER ATTEMPTED TO USE IT. THERE WAS NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INSTRUMENT WAS INSPECTED PRIOR TO USE AND LOOKED FINE. THE ISSUE OCCURRED WHILE THE SURGEON ATTEMPTED TO CLAMP ON A VESSEL OR TISSUE BUNDLE. THE PROCEDURE WAS COMPLETED ROBOTICALLY AFTER THEY OPENED ANOTHER CLIP APPLIER INSTRUMENT. NO FRAGMENTS FELL INSIDE THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
658045 ENDOWRIST LARGE HEM-O-LOK CLIP APPLIER NAY INTUITIVE SURGICAL, INC 470230-12 N10201027 0016 00886874112380

Patients

Seq Age Sex Outcome Treatment
1 NA Prefer Not To Disclose DA VINCI INSTRUMENTS AND ACCESSORIES