FDA Adverse Event Malfunction Summary report: N

SUNMED LLC

MDR report key: 19397864 · Received May 24, 2024

Report

Report Number
1314417-2024-00023
Event Type
Malfunction
Date Received
May 24, 2024
Date of Event
April 8, 2024
Report Date
June 27, 2024
Manufacturer
SUNMED HOLDINGS LLC
Product Code
BTM
UDI-DI
10888277628298
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

ALL INFORMATION REASONABLY KNOWN AS OF 24 MAY 2024 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AIRLIFE. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AIRLIFE. AIRLIFE. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AIRLIFE COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AIRLIFE PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Additional Manufacturer Narrative · 0

ALL INFORMATION REASONABLY KNOWN AS OF 24 MAY 2024 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AIRLIFE. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AIRLIFE. AIRLIFE. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AIRLIFE COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AIRLIFE PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY. THE COMPLAINT OF "BAG BODY BROKE AWAY FROM THE PATIENT PORT" REGARDING PART CPRM1116FM WAS CONFIRMED. THE ROOT CAUSE COULD POSSIBLY BE A RESULT OF THE PORT NOT BEING FULLY INSERTED INTO THE BAG DURING ASSEMBLY. A RISK ASSESSMENT WAS PERFORMED, AND THE ULTIMATE RISK WAS DETERMINED TO BE MEDIUM WHICH DOES REQUIRE REPORTING TO THE CAPA REVIEW BOARD. THERE HAVE BEEN 0 OTHER COMPLAINTS REGARDING THE SAME PART AND A SIMILAR ISSUE WITHIN THE 24 MONTHS PRECEDING THE REPORTING OF THIS ISSUE. A RESOLUTION LETTER WAS SENT TO THE CUSTOMER. COMPLAINTS WILL CONTINUE TO BE MONITORED FOR POSSIBLE TRENDS.

Description of Event or Problem · 0

RESUSCITATION BAGS THAT BROKE AWAY FROM THE BAG PART.

Description of Event or Problem · 0

RESUSCITATION BAGS THAT BROKE AWAY FROM THE BAG PART.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
643867 SUNMED LLC ADULT RESUS W/MASK, BAG RESERVOIR, FILTER & MANOMETER BTM SUNMED HOLDINGS LLC CPRM1116FM 461276 10888277628298

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other