SUNMED LLC
Report
- Report Number
- 1314417-2024-00023
- Event Type
- Malfunction
- Date Received
- May 24, 2024
- Date of Event
- April 8, 2024
- Report Date
- June 27, 2024
- Manufacturer
- SUNMED HOLDINGS LLC
- Product Code
- BTM
- UDI-DI
- 10888277628298
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- 501
Narratives
ALL INFORMATION REASONABLY KNOWN AS OF 24 MAY 2024 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AIRLIFE. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AIRLIFE. AIRLIFE. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AIRLIFE COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AIRLIFE PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.
ALL INFORMATION REASONABLY KNOWN AS OF 24 MAY 2024 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AIRLIFE. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AIRLIFE. AIRLIFE. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AIRLIFE COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AIRLIFE PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY. THE COMPLAINT OF "BAG BODY BROKE AWAY FROM THE PATIENT PORT" REGARDING PART CPRM1116FM WAS CONFIRMED. THE ROOT CAUSE COULD POSSIBLY BE A RESULT OF THE PORT NOT BEING FULLY INSERTED INTO THE BAG DURING ASSEMBLY. A RISK ASSESSMENT WAS PERFORMED, AND THE ULTIMATE RISK WAS DETERMINED TO BE MEDIUM WHICH DOES REQUIRE REPORTING TO THE CAPA REVIEW BOARD. THERE HAVE BEEN 0 OTHER COMPLAINTS REGARDING THE SAME PART AND A SIMILAR ISSUE WITHIN THE 24 MONTHS PRECEDING THE REPORTING OF THIS ISSUE. A RESOLUTION LETTER WAS SENT TO THE CUSTOMER. COMPLAINTS WILL CONTINUE TO BE MONITORED FOR POSSIBLE TRENDS.
RESUSCITATION BAGS THAT BROKE AWAY FROM THE BAG PART.
RESUSCITATION BAGS THAT BROKE AWAY FROM THE BAG PART.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 643867 | SUNMED LLC | ADULT RESUS W/MASK, BAG RESERVOIR, FILTER & MANOMETER | BTM | SUNMED HOLDINGS LLC | CPRM1116FM | 461276 | 10888277628298 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |