FDA Adverse Event Malfunction Summary report: N

ENDO GIA ULTRA

MDR report key: 19396315 · Received May 24, 2024

Report

Report Number
2647580-2024-02278
Event Type
Malfunction
Date Received
May 24, 2024
Date of Event
March 3, 2024
Report Date
May 24, 2024
Manufacturer
US SURGICAL PUERTO RICO
Product Code
GDW
UDI-DI
10884523003550
PMA / PMN Number
K111825
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D10 CONCOMITANT PRODUCT: EGIAUXL, EGIAUXL ENDOGIA ULTRA UNIV XL STAPLER, (LOT# P2H0242). 030419, 030419 ENDO GIAII 30 3.5MM DLU X6, (LOT# P2A0462). H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. NO INFORMATION RE GARDING THE SPECIFIC CUSTOMER ISSUE THAT OCCURRED WITH THE DEVICE WAS AVAILABLE. UPON EXAMINATION OF THE SAMPLE, SHEARED TEETH WERE VISIBLE ON THE FIRING RACK AT SITE OF INITIAL FIRING POST CLAMP UP. VISUAL INSPECTION NOTED THAT THE INSTRUMENT FIRING KNOBS WERE RETRACTED AND THE ARTICULATION LEVER WAS IN NEUTRAL POSITION. FUNCTIONALLY, THE INSTRUMENT WAS SUCCESSFULLY LOADED WITH A REPRESENTATIVE SINGLE USE LOADING UNIT (SULU). THE JAWS CLAMPED AND UNCLAMPED REPEATEDLY WITHOUT DIFFICULTY. THE INSTRUMENT WAS LOADED AND CLAMPED BUT WOULD NOT CYCLE DUE TO THE SHEARED FIRING RACK TEETH DAMAGE. AN ACCESS HOLE WAS CUT INTO THE INSTRUMENT BODY FOR VISUALIZATION OF THE FIRING RACK. THE PRODUCT ANALYSIS NOTED EVIDENCE THAT THE DEVICE WAS NOT USED AS INTENDED. SHEARED TEETH ON THE FIRING RACK CAN OCCUR IN ANY OF THE FOLLOWING CIRCUMSTANCES: FIRING OVER TISSUE THAT IS BEYOND THE RECOMMENDED THICKNESS RANGE, FIRING WITH AN OBSTACLE INCORPORATED IN THE JAWS OR ATTEMPTING TO FIRE A RELOAD THAT IS IN INTERLOCK. IN EACH CIRCUMSTANCE, IT WILL BECOME INCREASINGLY DIFFICULT TO ACTUATE THE FIRING HANDLE AND THE INSTRUMENT RETURN KNOBS WILL BE DIFFICULT TO RETRACT. IN ADDITION, STAPLES MAY NOT FORM PROPERLY, AND TISSUE MAY NOT BE FULLY TRANSECTED. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. THE INSTRUCTIONS INCLUDED WITH THIS DEVICE PROVIDE THE FOLLOWING GUIDANCE: PRE-OPERATIVE RADIOTHERAPY MAY RESULT IN CHANGES TO TISSUE. THESE CHANGES MAY, FOR EXAMPLE, CAUSE THE TISSUE THICKNESS TO EXCEED THE INDICATED RANGE THICKNESS FOR THE STAPLE SIZE. CAREFUL CONSIDERATION SHOULD BE GIVEN TO ANY PRE-SURGICAL TREATMENT THE PATIENT MAY HAVE UNDERGONE AND IN CORRESPONDING SELECTION OF STAPLE SIZE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC APPENDECTOMY, AFTER LOADING THE RELOAD, THE HANDLE WAS SQUEEZED, BUT THE JAWS DID NOT CLOSE AND RELEASED CORRECTLY. A NEW RELOAD WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT IS THAT THE SHEARED TEETH WERE VISIBLE ON THE FIRING RACK AT SITE OF INITIAL FIRING POST CLAMP UP. THE INSTRUMENT LOADED, CLAMPED, YET WOULD NOT CYCLE DUE TO THE SHEARED FIRING RACK TEETH DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2305696 ENDO GIA ULTRA STAPLE, IMPLANTABLE GDW US SURGICAL PUERTO RICO EGIAUXL P3C0336 10884523003550

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown