FDA Adverse Event Malfunction Summary report: N

BD PRESET¿

MDR report key: 19393299 · Received May 24, 2024

Report

Report Number
9617032-2024-00796
Event Type
Malfunction
Date Received
May 24, 2024
Date of Event
April 25, 2024
Report Date
May 6, 2024
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
50382903644150
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6. INVESTIGATION SUMMARY: "MATERIAL #: 364415. LOT/BATCH #: 3251818. BD HAD NOT RECEIVED SAMPLES, BUT 6 PHOTOS WERE PROVIDED FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR DEFORMED PLUNGER STOPPER WAS OBSERVED. ADDITIONALLY, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND THE ISSUE OF DEFORMED PLUNGER STOPPER WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE DEFORMED PLUNGER STOPPER. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS."

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING A BD PRESET¿ SYRINGE, THE PLUNGER STOPPER WAS DEFORMED WHICH CAUSED THE BLOOD COLLECTION TO FAIL. THE BLOOD SAMPLE NEEDED TO BE RECOLLECTED. NO IMPACT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
664450 BD PRESET¿ BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 3251818 50382903644150

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown