BD PRESET¿
Report
- Report Number
- 9617032-2024-00796
- Event Type
- Malfunction
- Date Received
- May 24, 2024
- Date of Event
- April 25, 2024
- Report Date
- May 6, 2024
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- UDI-DI
- 50382903644150
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6. INVESTIGATION SUMMARY: "MATERIAL #: 364415. LOT/BATCH #: 3251818. BD HAD NOT RECEIVED SAMPLES, BUT 6 PHOTOS WERE PROVIDED FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR DEFORMED PLUNGER STOPPER WAS OBSERVED. ADDITIONALLY, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND THE ISSUE OF DEFORMED PLUNGER STOPPER WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE DEFORMED PLUNGER STOPPER. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS."
IT WAS REPORTED WHILE USING A BD PRESET¿ SYRINGE, THE PLUNGER STOPPER WAS DEFORMED WHICH CAUSED THE BLOOD COLLECTION TO FAIL. THE BLOOD SAMPLE NEEDED TO BE RECOLLECTED. NO IMPACT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 664450 | BD PRESET¿ | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON AND COMPANY (BD) | 3251818 | 50382903644150 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |