FDA Adverse Event Malfunction Summary report: N

ZEVO¿ ANTERIOR CERVICAL PLATE SYSTEM

MDR report key: 19389595 · Received May 24, 2024

Report

Report Number
1030489-2024-00550
Event Type
Malfunction
Date Received
May 24, 2024
Date of Event
March 15, 2024
Report Date
May 27, 2024
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC
Product Code
HXX
UDI-DI
00643169342293
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS #PE 706377581, PART # 3030007, LOT # IM18C018 VISUAL INSPECTION REVEALED THE TIP OF THE SCREWDRIVER HAS BEEN BROKEN OFF. DAMAGE PRESENT AT BREAK POINT IS CONSISTENT WITH TORSIONAL OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

G2: REPORT SOURCE UPDATED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A SCREWDRIVER USED FOR SPINAL THERAPY. IT WAS REPORTED THAT THE DEVICE HAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2273634 ZEVO¿ ANTERIOR CERVICAL PLATE SYSTEM SCREWDRIVER HXX MEDTRONIC SOFAMOR DANEK USA, INC 3030007 IM18C018 00643169342293

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown