FDA Adverse Event Injury Summary report: N

DAVINCI XI

MDR report key: 19389090 · Received May 23, 2024

Report

Report Number
2955842-2024-14992
Event Type
Injury
Date Received
May 23, 2024
Date of Event
April 23, 2024
Report Date
April 23, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SECTION D10 CONCOMITANT PRODUCTS: ERBE VIO DV 2.0 INTEGRATED ELECTROSURGICAL UNIT, PRODUCT MATERIAL # 951300-03, SERIAL NUMBER (B)(6). THE ERBE GENERATOR (OEM) IS INTEGRATED WITH THE DAVINCI VISION SIDE CART. THE PROCEDURE WAS AN OPEN APPENDECTOMY AND NO OTHER DAVINCI ROBOTIC CARTS OR PRODUCTS WERE USED. THE ERBE IESU GENERATOR WAS NOT MADE AVAILABLE TO THE FIELD SERVICE ENGINEER FOR EVALUATION. THE DAVINCI VISION SIDE CART WITH THE NEWLY INSTALLED ERBE GENERATOR PASSED TESTING AND WAS VERIFIED AS READY FOR USE.

Description of Event or Problem · 0

IT WAS REPORTED THAT A DAVINCI VISION SIDE CART'S ERBE VIO DV 2.0 INTEGRATED ELECTROSURGICAL UNIT (ERBE IESU) WAS USED DURING A NON-ROBOTIC, OPEN APPENDECTOMY PROCEDURE. AN OR NURSE INFORMED INTUITIVE SURGICAL TECHNICAL SUPPORT THAT DURING THE OPEN PROCEDURE "THEY HAD A POSSIBLE BURN WITH THE ERBE IESU...IT MAY HAVE BURNED THE PATIENT." ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION FROM THE CUSTOMER SITE HAVE NOT BEEN SUCCESSFUL. AN INTUITIVE SURGICAL FIELD SERVICE ENGINEER (FSE) FOUND THAT THE ERBE HAD ALREADY BEEN REMOVED FROM THE DAVINCI VISION SIDE CART. THE UNIT WAS SECURED IN THE BIOMEDICAL ENGINEERING DEPARTMENT AND WAS WAS NOT MADE AVAILABLE FOR THE FSE TO PERFORM SYSTEM VERIFICATION AND ELECTRICAL SAFETY EVALUATIONS. A NEW ERBE IESU WAS INSTALLED IN THE VISION SIDE CART AND THE UNIT PASSED SYSTEM VERIFICATION AND ELECTRICAL SAFETY TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
664291 DAVINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-45 00886874110898

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention DA VINCI INSTRUMENTS AND ACCESSORIES