FDA Adverse Event Malfunction Summary report: N

3M¿ RANGER¿ BLOOD/FLUID WARMING HIGH FLOW SET

MDR report key: 19388144 · Received May 23, 2024

Report

Report Number
2110898-2024-00021
Event Type
Malfunction
Date Received
May 23, 2024
Date of Event
April 22, 2024
Report Date
May 23, 2024
Manufacturer
3M COMPANY
Product Code
BSB
UDI-DI
10707387792685
PMA / PMN Number
K973741
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT BEEN RETURNED TO SOLVENTUM FOR ANALYSIS. SOLVENTUM IS UNABLE TO VERIFY THE LEAK, IDENTIFY EXACT LOCATION AND ROOT CAUSE WITHOUT THE SAMPLE. BASED ON THE PROBLEM DESCRIPTION AND PHOTO PROVIDED, A LIKELY CAUSE IS A HEAT EXCHANGER LEAK. POSSIBLE CAUSES OF HEAT EXCHANGER LEAKS INCLUDE OVER-PRESSURIZATION ABOVE 300MMHG, INSERTION OR REMOVAL OF PRESSURIZED HEAT EXCHANGER FROM HEATING UNIT, AND HEAT EXCHANGER NOT FULLY INSERTED INTO THE HEATING UNIT. A REVIEW OF THE BATCH RECORD SHOWED THIS LOT MET ALL SPECIFICATION FOR PRODUCT RELEASE AND NO DEVIATIONS WERE FOUND THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED MALFUNCTION. THERE IS NO CHANGE IN TRENDING FOR HEAT EXCHANGER LEAKS. SOLVENTUM WILL CONTINUE TO MONITOR.

Additional Manufacturer Narrative · 0

MDR 2110898-2024-00021 SUBMITTED ON 23-MAY-2024 NOTED THE FOLLOWING: SECTION D4 UNIQUE IDENTIFIER (UDI) #: (B)(4). CORRECTION: SECTION D4 UNIQUE IDENTIFIER (UDI) #: (B)(4). THE DEVICE WAS RETURNED TO SOLVENTUM FOR ANALYSIS. THE SAMPLE CONFIRMED THERE WAS A LEAK AT THE HEAT EXCHANGER. CAUSE AND TIMING OF THE LEAK COULD NOT BE DETERMINED. SOLVENTUM WILL CONTINUE TO MONITOR.

Description of Event or Problem · 0

THE USER FACILITY REPORTED THAT THE 3M¿ RANGER¿ BLOOD/FLUID WARMING HIGH FLOW SET LEAKED AT THE PROXIMAL END OF THE CASSETTE. NO INJURIES OR MEDICAL INTERVENTIONS WERE REPORTED DUE TO THE ALLEGED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2244791 3M¿ RANGER¿ BLOOD/FLUID WARMING HIGH FLOW SET WARMER, BLOOD, NONELECTROMAGNETIC RADIATION BSB 3M COMPANY 24355 HX9304 10707387792685

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown