FDA Adverse Event Malfunction Summary report: N

DRX REVOLUTION MOBILE X-RAY SYSTEM

MDR report key: 19381839 · Received May 23, 2024

Report

Report Number
1317307-2024-00006
Event Type
Malfunction
Date Received
May 23, 2024
Date of Event
April 24, 2024
Report Date
June 5, 2024
Manufacturer
CARESTREAM HEALTH INC.
Product Code
IZL
UDI-DI
60889978618897
PMA / PMN Number
K191025
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CARESTREAM HEALTH INC. (CSH) HAS INVESTIGATED THIS ISSUE AND IT WAS DETERMINED THAT THIS INCIDENT WAS DUE TO A DOCK MECHANISM ISSUE. THE REVOLUTION SYSTEM IS DESIGNED TO DRIVE AT VERY LOW SPEED WHEN THE TUBE COLUMN IS UNDOCKED. IN THIS CASE, THE DOCKING MECHANISM WAS INCORRECTLY PROVIDING A SIGNAL OF "UNDOCKED CONDITION" WHILE DRIVING, CAUSING THE UNIT TO COME TO A STOP AS DESIGNED. ALTHOUGH THE UNIT CONTACTED THE PATIENT, IT DID NOT RESULT IN AN INJURY. THE LATCH ROTARY ASSEMBLY AND THE CABLE-MAC TO BOOM LOCK AND COLUMN DOWN SENSOR WERE REPLACED. BREAK-FIX. THE SYSTEM IS CURRENTLY IN USE WITH NO RE-OCCURENCES. THIS TYPE OF RISK IS ALREADY INCLUDED IN THE RISK ASSESSMENT AND CONTROL MATRIX OF THE PRODUCT. THE RISK HAS BEEN MITIGATED AS FAR AS POSSIBLE. THERE ARE NO FURTHER ACTIONS TO BE TAKEN BY CARESTREAM HEALTH INC. RELATED TO THIS ISSUE. CARESTREAM HAS COMPLETED THIS INVESTIGATION.

Additional Manufacturer Narrative · 0

CSH INVESTIGATION IS CURRENTLY IN PROGRESS. SUBMITTING AN INITIAL REPORT AS THERE ARE CURRENTLY INSUFFICIENT FACTS AVAILABLE TO MAKE A REPORTABILITY DECISION AND AT LEAST 28 DAYS HAVE ELAPSED. CSH WILL SUBMIT A FOLLOW UP WITH INVESTIGATION UPDATE/FINDINGS WITHIN 30 DAYS.

Description of Event or Problem · 0

ON (B)(6) 2024, CARESTREAM HEALTH (CSH) WAS INFORMED OF WAS INFORMED OF AN UNINTENDED MOVEMENT WHILE USING THE DRX-REVOLUTION PLUS MOBILE X-RAY SYSTEM. NO INJURIES REPORTED.

Description of Event or Problem · 0

ON (B)(6) 2024, CARESTREAM HEALTH (CSH) WAS INFORMED OF WAS INFORMED OF AN UNINTENDED MOVEMENT WHILE USING THE DRX-REVOLUTION PLUS MOBILE X-RAY SYSTEM. NO INJURIES REPORTED, THE INVESTIGATION IS ONGOING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1616478 DRX REVOLUTION MOBILE X-RAY SYSTEM DRX REVOLUTION MOBILE X-RAY SYSTEM IZL CARESTREAM HEALTH INC. 60889978618897

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other