FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER
MDR report key: 1937281
·
Received December 23, 2010
Report
- Report Number
- 2050012-2010-01670
- Event Type
- Malfunction
- Date Received
- December 23, 2010
- Date of Event
- November 30, 2010
- Report Date
- November 30, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K060256
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER IS UNSURE OF WHAT FLUID IS LEAKING. A BCI FIELD SERVICE ENGINEER (FSE) FOUND THAT THE WASH BUFFER WAS GETTING INTO AUTOGLOSS WELL. FSE PERFORMED THE FOLLOWINGS: REPLACED THE PERIPUMP TUBES CLEANED AUTOGLOSS WELL. RE-TAPED WASH BUFFER CONNECTOR ON RESERVOIR WHICH WAS LEAKING. PRIMED THE SYSTEM THE ISSUE HAS BEEN FIXED.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO CLOSE TUBE ACCESSION (CTA) LEAK. NO INJURY AND/OR PERSONNEL EXPOSURE TO CHEMICAL OR BIO-HAZARDOUS MATERIAL WAS REPORTED.