FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 1937281 · Received December 23, 2010

Report

Report Number
2050012-2010-01670
Event Type
Malfunction
Date Received
December 23, 2010
Date of Event
November 30, 2010
Report Date
November 30, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K060256
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER IS UNSURE OF WHAT FLUID IS LEAKING. A BCI FIELD SERVICE ENGINEER (FSE) FOUND THAT THE WASH BUFFER WAS GETTING INTO AUTOGLOSS WELL. FSE PERFORMED THE FOLLOWINGS: REPLACED THE PERIPUMP TUBES CLEANED AUTOGLOSS WELL. RE-TAPED WASH BUFFER CONNECTOR ON RESERVOIR WHICH WAS LEAKING. PRIMED THE SYSTEM THE ISSUE HAS BEEN FIXED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO CLOSE TUBE ACCESSION (CTA) LEAK. NO INJURY AND/OR PERSONNEL EXPOSURE TO CHEMICAL OR BIO-HAZARDOUS MATERIAL WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 600I N/A

Patients

Seq Age Sex Outcome Treatment
1