CONCERTO
Report
- Report Number
- 9710014-2024-00454
- Event Type
- Malfunction
- Date Received
- May 22, 2024
- Date of Event
- May 19, 2024
- Report Date
- January 30, 2025
- Manufacturer
- MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
- Product Code
- MCM
- UDI-DI
- 09008737306588
- PMA / PMN Number
- P000025
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SA
- Reporter Occupation
- 003
Narratives
CONCLUSION: OVERLOAD FRACTURES IN THE ACTIVE ELECTRODE WHICH ARE CONSISTENT WITH AN EXTERNAL MECHANICAL IMPACT WERE DETERMINED TO BE THE ROOT CAUSE OF DEVICE FAILURE. THE ACTIVE ELECTRODE ALSO SHOWED SOME ADDITIONAL DAMAGE TO THE ELECTRODE WIRES, AS BEING CAUSED BY MINUTE DEVICE MOBILITY. AN ACCIDENT OR IMPACT MIGHT HAVE WEAKENED THE FIXATION OF THE IMPLANT, CONSEQUENTLY LEADING TO ELECTRODE MOBILITY. THE PROBLEMS GIVEN IN THE RECIPIENT REPORT APPEAR TO MATCH THE DAMAGE FOUND. THIS IS A FINAL REPORT.
THE DEVICE HAS NOT BEEN EXPLANTED. IF IT SHOULD BE EXPLANTED, IT SHOULD BE RETURNED TO THE MANUFACTURER FOR EVALUATION. WHEN AVAILABLE, A DEVICE FAILURE ANALYSIS WILL BE SUBMITTED AS A FOLLOW UP REPORT.
ADDITIONAL INFORMATION: ACCORDING TO THE CURRENTLY AVAILABLE INFORMATION, DAMAGE TO THE ACTIVE ELECTRODE LIKELY CAUSED BY MINUTE DEVICE MOBILITY IS SUSPECTED. HOWEVER, TO DETERMINE AN EXACT ROOT CAUSE A DEVICE INVESTIGATION OF THE EXPLANTED DEVICE IS NECESSARY. RE-IMPLANTATION IS CONSIDERED BUT NO DATE HAS BEEN SCHEDULED YET.
THE USER'S HEARING PERFORMANCE GRADUALLY WORSENED. THE PARENTS REPORTED THAT THE USER MIGHT HAVE HIT HER HEAD WHILE BATHING. THE USER HAS BEEN RE-IMPLANTED.
THE USER'S HEARING PERFORMANCE GRADUALLY WORSENED. THE PARENTS REPORTED THAT THE USER MIGHT HAVE HIT HER HEAD WHILE BATHING. REIMPLANTATION IS BEING CONSIDERED, BUT NO DATE HAS BEEN SCHEDULED YET.
THE USER'S HEARING PERFORMANCE GRADUALLY WORSENED. THE PARENTS REPORTED THAT THE USER MIGHT HAVE HIT HER HEAD WHILE BATHING. REIMPLANTATION IS BEING CONSIDERED, BUT NO DATE HAS BEEN SCHEDULED YET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1748954 | CONCERTO | COCHLEAR IMPLANT | MCM | MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH | MI1000 MED-EL CONCERT | 09008737306588 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR | Female | Required Intervention |