FDA Adverse Event Malfunction Summary report: N

CONCERTO

MDR report key: 19368920 · Received May 22, 2024

Report

Report Number
9710014-2024-00454
Event Type
Malfunction
Date Received
May 22, 2024
Date of Event
May 19, 2024
Report Date
January 30, 2025
Manufacturer
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
Product Code
MCM
UDI-DI
09008737306588
PMA / PMN Number
P000025
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CONCLUSION: OVERLOAD FRACTURES IN THE ACTIVE ELECTRODE WHICH ARE CONSISTENT WITH AN EXTERNAL MECHANICAL IMPACT WERE DETERMINED TO BE THE ROOT CAUSE OF DEVICE FAILURE. THE ACTIVE ELECTRODE ALSO SHOWED SOME ADDITIONAL DAMAGE TO THE ELECTRODE WIRES, AS BEING CAUSED BY MINUTE DEVICE MOBILITY. AN ACCIDENT OR IMPACT MIGHT HAVE WEAKENED THE FIXATION OF THE IMPLANT, CONSEQUENTLY LEADING TO ELECTRODE MOBILITY. THE PROBLEMS GIVEN IN THE RECIPIENT REPORT APPEAR TO MATCH THE DAMAGE FOUND. THIS IS A FINAL REPORT.

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT BEEN EXPLANTED. IF IT SHOULD BE EXPLANTED, IT SHOULD BE RETURNED TO THE MANUFACTURER FOR EVALUATION. WHEN AVAILABLE, A DEVICE FAILURE ANALYSIS WILL BE SUBMITTED AS A FOLLOW UP REPORT.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: ACCORDING TO THE CURRENTLY AVAILABLE INFORMATION, DAMAGE TO THE ACTIVE ELECTRODE LIKELY CAUSED BY MINUTE DEVICE MOBILITY IS SUSPECTED. HOWEVER, TO DETERMINE AN EXACT ROOT CAUSE A DEVICE INVESTIGATION OF THE EXPLANTED DEVICE IS NECESSARY. RE-IMPLANTATION IS CONSIDERED BUT NO DATE HAS BEEN SCHEDULED YET.

Description of Event or Problem · 0

THE USER'S HEARING PERFORMANCE GRADUALLY WORSENED. THE PARENTS REPORTED THAT THE USER MIGHT HAVE HIT HER HEAD WHILE BATHING. THE USER HAS BEEN RE-IMPLANTED.

Description of Event or Problem · 0

THE USER'S HEARING PERFORMANCE GRADUALLY WORSENED. THE PARENTS REPORTED THAT THE USER MIGHT HAVE HIT HER HEAD WHILE BATHING. REIMPLANTATION IS BEING CONSIDERED, BUT NO DATE HAS BEEN SCHEDULED YET.

Description of Event or Problem · 0

THE USER'S HEARING PERFORMANCE GRADUALLY WORSENED. THE PARENTS REPORTED THAT THE USER MIGHT HAVE HIT HER HEAD WHILE BATHING. REIMPLANTATION IS BEING CONSIDERED, BUT NO DATE HAS BEEN SCHEDULED YET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1748954 CONCERTO COCHLEAR IMPLANT MCM MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MI1000 MED-EL CONCERT 09008737306588

Patients

Seq Age Sex Outcome Treatment
1 13 YR Female Required Intervention