FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 19368452 · Received May 22, 2024

Report

Report Number
9610877-2024-53730
Event Type
Malfunction
Date Received
May 22, 2024
Date of Event
May 14, 2024
Report Date
May 22, 2024
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDT
UDI-DI
04961333233007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD DRIVE PCB AS DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD DRIVE PCB. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER TUBE CLOGGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1749951 PENTAX VIDEO DUODENOSCOPE FDT HOYA CORPORATION PENTAX TOKYO OFFICE ED34-I10T2 04961333233007

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown