FDA Adverse Event Malfunction Summary report: N

COLLEAGUE VOLUMETRIC INFUSOR PUMP-SINGLE CHANNEL

MDR report key: 1936571 · Received December 22, 2010

Report

Report Number
6000001-2010-06173
Event Type
Malfunction
Date Received
December 22, 2010
Date of Event
September 1, 2010
Report Date
September 20, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA-(B)(4). A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

THE REPORTED CONDITION OF FAILURE CODE 812:02 WAS CONFIRMED BUT NOT DUPLICATED. THE REPORTED CONDITION IS DUE TO A FAULTY PHM (PUMP HEAD MODULE). THE PHM WAS REPLACED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4)

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 812:02. THE EVENT WAS REPORTED TO HAVE OCCURRED UPON POWER UP. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 5.09.90 CATEGORIZED AS REMEDIATED. NO ADDITIONAL INFORMATION IS AVAILABLE. DURING REVIEW OF THE EVENT HISTORY BY BAXTER IT WAS DETERMINED THAT THE REPORTED CONDITION OCCURRED DURING DELIVERY CAUSING AN INTERRUPTION OF DELIVERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE VOLUMETRIC INFUSOR PUMP-SINGLE CHANNEL PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1