BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)
Report
- Report Number
- 2647876-2024-00078
- Event Type
- Injury
- Date Received
- May 21, 2024
- Date of Event
- April 22, 2024
- Report Date
- June 17, 2024
- Manufacturer
- BECTON DICKINSON CARIBE LTD.
- Product Code
- MDB
- UDI-DI
- 00382904420239
- PMA / PMN Number
- K113558
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6. INVESTIGATION SUMMARY: CUSTOMER REPORTED A CONTAMINATION ISSUE WHILE USING BACTEC PRODUCT. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH BACTEC PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS.
G5: PMA/510(K)#: K222591. H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHEN USING THE BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC), THERE WAS ONE CONTAMINATED BLOOD CULTURE BOTTLE. THE CONTAMINATED CULTURE GREW STAPH SAPRO. UNAWARE OF THE TREATMENT ADMINISTERED AT THIS TIME. NO ADVERSE IMPACT REPORTED.
IT WAS REPORTED THAT WHEN USING THE BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC), THERE WAS ONE CONTAMINATED BLOOD CULTURE BOTTLE. THE CONTAMINATED CULTURE GREW STAPH SAPRO. UNAWARE OF THE TREATMENT ADMINISTERED AT THIS TIME. NO ADVERSE IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 515943 | BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) | SYSTEM, BLOOD CULTURING | MDB | BECTON DICKINSON CARIBE LTD. | 3284978 | 00382904420239 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |