FDA Adverse Event Injury Summary report: N

BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 19363247 · Received May 21, 2024

Report

Report Number
2647876-2024-00077
Event Type
Injury
Date Received
May 21, 2024
Date of Event
April 22, 2024
Report Date
August 21, 2024
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420239
PMA / PMN Number
K113558
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVALUATION: YES. H.3 DEVICE EVAL BY MANUFACTURER? YES. D9: RETURNED TO MANUFACTURER ON: 06-MAY-2024. INVESTIGATION SUMMARY: CATALOG 442023. BATCH NO. 3340577. CUSTOMER STATES THEY ARE RECEIVING STRANGE POSITIVE BLOOD CULTURES. BD WAS UNABLE TO REPRODUCE CUSTOMER EXPERIENCE WITH THE BACTEC PRODUCT. A FALSE POSITIVE RESPONSE WAS NOT OBSERVED WHEN RETENTION/RETURNED SAMPLES WERE TESTED. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORDS WERE REVIEWED, AND ALL TESTING WERE WITHIN SPECIFICATION FOR PRODUCT RELEASE. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. COMPLAINT IS UNCONFIRMED. NO CORRECTIVES ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATION PROCESS.

Additional Manufacturer Narrative · 0

G5: PMA/510(K)#: K222591. H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

REPORT 2 OF 2. IT WAS REPORTED THAT WHEN USING THE BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC), THERE WAS ONE CONTAMINATED BLOOD CULTURE BOTTLE. THE CONTAMINATED CULTURE GREW PANTOEA AGGLOMERANS, PSEUDOMONAS SPP, AND PSEUDOOMONAS STUZI. UNAWARE OF THE TREATMENT ADMINISTERED AT THIS TIME. NO ADVERSE IMPACT REPORTED.

Description of Event or Problem · 0

REPORT 2 OF 2. IT WAS REPORTED THAT WHEN USING THE BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC), THERE WAS ONE CONTAMINATED BLOOD CULTURE BOTTLE. THE CONTAMINATED CULTURE GREW PANTOEA AGGLOMERANS, PSEUDOMONAS SPP, AND PSEUDOOMONAS STUZI. UNAWARE OF THE TREATMENT ADMINISTERED AT THIS TIME. NO ADVERSE IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1155542 BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 3340577 00382904420239

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other