FDA Adverse Event Malfunction Summary report: N

HOMECHOICE CYCLER-REFURBISHED

MDR report key: 1936292 · Received December 22, 2010

Report

Report Number
1423500-2010-07167
Event Type
Malfunction
Date Received
December 22, 2010
Date of Event
November 13, 2010
Report Date
December 2, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED DIFFICULTY. THE DEVICE MET SPECIFICATIONS. THE ROOT CAUSE WAS DETERMINED TO BE INSUFFICIENT DRAIN: ONE OR MORE CYCLES ADVANCES TO THE NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MIN DRAIN VOLUME THRESHOLD. REVIEW OF THE SERVICE DATES AND ACTIVITIES REVEALED THE PREVIOUS RETURN OF THE DEVICE WAS NOT FOR THE REPORTED PROBLEMS OF IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE IIPV OCCURRED ON (B)(6) 2010 DURING DRAIN CYCLE 4. THE PROGRAMMED FILL VOLUME WAS 1200ML AND THE TOTAL DRAIN VOLUME WAS 2007ML. THIS DRAIN VOLUME MEETS BAXTER'S IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE CYCLER-REFURBISHED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 67 YR