ENDOPATH** XCEL*
Report
- Report Number
- 3005075853-2010-07348
- Event Type
- Malfunction
- Date Received
- December 22, 2010
- Date of Event
- December 6, 2010
- Report Date
- December 9, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GCJ
- PMA / PMN Number
- K032676
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. UPON EVALUATION OF THE DEVICE, THE DUCKBILL WAS FOUND TO BE SLIGHTLY OPEN AT SLIT. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK DURING INSERTION AND REMOVAL OF THE TEST PROBE THROUGH THE DEVICE. A POTENTIAL CAUSE OF THIS FAILURE IS ATTRIBUTED TO MOLDING, COMPONENT TRANSIT, BODILY FLUIDS OR DEBRIS FROM SURGERY. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. HOWEVER, AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE POTENTIAL ROOT CAUSE OF THE INSUFFLATIONS ISSUES. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, AIR LEAKED A LOT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. PLEASE TAKE PHOTOS FOR (B)(6) CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH** XCEL* | LAPAROSCOPE, GENERAL AND PLASTIC SURGERY | GCJ | ETHICON ENDO-SURGERY, LLC. | NA | G4TJ35 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |