FDA Adverse Event Malfunction Summary report: N

ENDOPATH** XCEL*

MDR report key: 1936111 · Received December 22, 2010

Report

Report Number
3005075853-2010-07345
Event Type
Malfunction
Date Received
December 22, 2010
Date of Event
December 6, 2010
Report Date
December 6, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS FUNCTIONALLY TESTED TO DETECT ANY LEAKING ISSUES. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK WITH THE TEST PROBE INSERTED THROUGH THE DEVICE. IT SHOULD BE NOTED THAT AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE POTENTIAL ROOT CAUSE OF THE REPORTED INSUFFLATIONS ISSUES. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. THREE ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. (B)(4)

Additional Manufacturer Narrative · 1

(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ENDOSCOPIC THYROIDECTOMY PROCEDURE, THERE WAS A GAS LEAK. UNKNOWN HOW CASE WAS COMPLETED. SURGERY WAS PROLONGED FIVE MINUTES. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** XCEL* LAPAROSCOPE, GENERAL AND PLASTIC SURGERY GCJ ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1