FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 1936045 · Received December 22, 2010

Report

Report Number
2953144-2010-03340
Event Type
Injury
Date Received
December 22, 2010
Date of Event
November 29, 2010
Report Date
November 29, 2010
Manufacturer
AV-REDWOOD CITY
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: EVALUATION OF THE RETURNED DEVICE FOUND THAT THE PLUNGER, CUFFS, LINK, NEEDLE TIP AND MONOFILAMENT WERE NOT RETURNED. IT WAS REPORTED THAT THE KNOT WAS FORMED OUTSIDE THE BODY. THE PROBABLE CAUSE FOR THIS CONDITION IS THE DEVICE WAS DEPLOYED OUTSIDE THE ARTERY AND THE KNOT LOCKED WHEN THE NON-RAIL END OF THE MONOFILAMENT WAS PULLED INADVERTENTLY. SINCE THE MONOFILAMENT WAS NOT RETURNED, WHICH IS AN INTEGRAL PART OF THIS INVESTIGATION, IT WAS NOT POSSIBLE TO DETERMINE THE CAUSE OR CONFIRM THE COMPLAINT. NO MANUFACTURING OR QUALITY ISSUE WAS DETECTED. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4): DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF A MILDLY CALCIFIED COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, DURING SUTURE HARVESTING, THE KNOT WAS FORMED 2-3 INCHES OUTSIDE THE PATIENT'S ANATOMY AND,THEREFORE, COULD NOT BE ADVANCED. THE SUTURES WERE REMOVED AND A NON-ABBOTT DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THOUGH REQUESTED, ADDITIONAL INFORMATION WAS NOT PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-REDWOOD CITY 940016H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention