FDA Adverse Event Injury Summary report: N

CANALIZER STANDARD ANGLED

MDR report key: 1936030 · Received December 9, 2010

Report

Report Number
9613246-2010-00003
Event Type
Injury
Date Received
December 9, 2010
Report Date
December 3, 2010
Manufacturer
PBN MEDICALS DENMARK A/S
Product Code
DQX
PMA / PMN Number
K050873
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

EVAL OF THE REPORTED PRODUCT CONFIRMED THE CUSTOMER'S REPORT. THERE WERE NO DEFECTS OR DEVIATIONS NOTED DURING THE BATCH RECORD REVIEW. BASED ON THE EVAL OF THE RETURNED PRODUCT WE CONCLUDE THAT THE POTENTIAL REASON FOR THE DAMAGE TO THE PRODUCT IS A RESULT OF INCORRECT USE, POTENTIALLY FROM CONTACT WITH A SHARP DEVICE, E.G. AN INTRODUCER NEEDLE. OUR IFU STATES: "CAUTION: DO NOT USE THE GUIDEWIRE WITH DEVICES THAT CONTAIN METAL PARTS, SUCH AS ATHERECTOMY CATHETER OR METAL INTRODUCING DEVICES. USE WITH DEVICES THAT CONTAIN METAL PART MAY CAUSE DAMAGE AND/OR ABRASION OF THE HYDROPHILIC COATING. TO PREVENT DAMAGE TO THE VASCULATURE AND GUIDEWIRE AND BREAKAGE/SEPARATION/CUTTING OF THE CATHETER, ALWAYS MANIPULATE CAREFULLY AND SLOWLY WHILE CONFIRMING THE MOTION AND LOCATION OF THE GUIDEWIRE TIP UNDER FLUOROSCOPY. DURING MANIPULATION, IF ANY RESISTANCE IS FELT OR IF THERE APPEARS TO BE ANY PROBLEMS WITH THE GUIDEWIRES TIP, STOP THE MANIPULATION AND FIND OUT THE CAUSE WITH THE FLUOROSCOPY. CONTINUING A MANIPULATION AFTER A PROBLEM HAS BEEN DETECTED MIGHT CAUSE DAMAGE TO THE VASCULATURE, BREAKAGE/SEPARATION/CUTTING OF THE GUIDEWIRE AND/OR DAMAGE TO THE CATHETER."

Description of Event or Problem · 1

LIMITED INFO OBTAINED FROM COMPLAINANT. DURING A VASCULAR PROCEDURE USING A CANALIZER GUIDE WIRE PRODUCT, CUSTOMER STATES THAT THE POLYURETHANE JACKET OF THE GUIDEWIRE CAME OFF APPROX 10-15CM OF THE POLYURETHANE JACKET WAS LEFT IN THE PT. AN ADDITIONAL PROCEDURE WAS REQUIRED TO REMOVE THE FRAGMENT FROM THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CANALIZER STANDARD ANGLED CANALIZER GUIDE WIRE DQX PBN MEDICALS DENMARK A/S 127635180 P353527

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention