CANALIZER STANDARD ANGLED
Report
- Report Number
- 9613246-2010-00003
- Event Type
- Injury
- Date Received
- December 9, 2010
- Report Date
- December 3, 2010
- Manufacturer
- PBN MEDICALS DENMARK A/S
- Product Code
- DQX
- PMA / PMN Number
- K050873
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
EVAL OF THE REPORTED PRODUCT CONFIRMED THE CUSTOMER'S REPORT. THERE WERE NO DEFECTS OR DEVIATIONS NOTED DURING THE BATCH RECORD REVIEW. BASED ON THE EVAL OF THE RETURNED PRODUCT WE CONCLUDE THAT THE POTENTIAL REASON FOR THE DAMAGE TO THE PRODUCT IS A RESULT OF INCORRECT USE, POTENTIALLY FROM CONTACT WITH A SHARP DEVICE, E.G. AN INTRODUCER NEEDLE. OUR IFU STATES: "CAUTION: DO NOT USE THE GUIDEWIRE WITH DEVICES THAT CONTAIN METAL PARTS, SUCH AS ATHERECTOMY CATHETER OR METAL INTRODUCING DEVICES. USE WITH DEVICES THAT CONTAIN METAL PART MAY CAUSE DAMAGE AND/OR ABRASION OF THE HYDROPHILIC COATING. TO PREVENT DAMAGE TO THE VASCULATURE AND GUIDEWIRE AND BREAKAGE/SEPARATION/CUTTING OF THE CATHETER, ALWAYS MANIPULATE CAREFULLY AND SLOWLY WHILE CONFIRMING THE MOTION AND LOCATION OF THE GUIDEWIRE TIP UNDER FLUOROSCOPY. DURING MANIPULATION, IF ANY RESISTANCE IS FELT OR IF THERE APPEARS TO BE ANY PROBLEMS WITH THE GUIDEWIRES TIP, STOP THE MANIPULATION AND FIND OUT THE CAUSE WITH THE FLUOROSCOPY. CONTINUING A MANIPULATION AFTER A PROBLEM HAS BEEN DETECTED MIGHT CAUSE DAMAGE TO THE VASCULATURE, BREAKAGE/SEPARATION/CUTTING OF THE GUIDEWIRE AND/OR DAMAGE TO THE CATHETER."
LIMITED INFO OBTAINED FROM COMPLAINANT. DURING A VASCULAR PROCEDURE USING A CANALIZER GUIDE WIRE PRODUCT, CUSTOMER STATES THAT THE POLYURETHANE JACKET OF THE GUIDEWIRE CAME OFF APPROX 10-15CM OF THE POLYURETHANE JACKET WAS LEFT IN THE PT. AN ADDITIONAL PROCEDURE WAS REQUIRED TO REMOVE THE FRAGMENT FROM THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CANALIZER STANDARD ANGLED | CANALIZER GUIDE WIRE | DQX | PBN MEDICALS DENMARK A/S | 127635180 | P353527 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |