HOMECHOICE CYCLER-REFURBISHED
Report
- Report Number
- 1423500-2010-07151
- Event Type
- Malfunction
- Date Received
- December 22, 2010
- Date of Event
- October 10, 2010
- Report Date
- December 1, 2010
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE RITE (RETURN INSTRUMENT TEST/EVALUATION) TEST WAS PERFORMED WHEN THE DEVICE WAS RETURNED TO THE BAXTER TAMPA BAY FACILITY FOR EVALUATION. THE DEVICE PASSED THE HOMECHOICE RITE FUNCTIONAL TEST AND THE HOMECHOICE RITE ELECTRICAL TEST. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE INSUFFICIENT DRAIN, ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). DEVICE EVALUATION EXPECTED, BUT NOT YET COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF BAXTER'S INVESTIGATION.
AN INSTANCE OF INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED DURING A REVIEW OF THE RETURNED HOMECHOICE (HC) PATIENT EVENT LOG. ON (B)(6) 2010, THE DRAIN VOLUME WAS 3289ML DURING CYCLE 6.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE CYCLER-REFURBISHED | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |