FDA Adverse Event Malfunction Summary report: N

HEARTSTRING III PROXIMAL SEAL SYSTEM, 4.3 MM

MDR report key: 1935852 · Received December 3, 2010

Report

Report Number
2242352-2010-03520
Event Type
Malfunction
Date Received
December 3, 2010
Date of Event
November 4, 2010
Report Date
November 5, 2010
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED WITHOUT THE LOADING DEVICE. THE TENSION SPRING ASSEMBLY AND THE SEAL WERE RETURNED SEPARATE. THE SEAL WAS INTACT. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE DEPRESSED ON THE DELIVERY DEVICE. THE DEVICE WAS VERY BLOODY. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL FELL APART WHEN PULLING OUT OF AORTA" WAS CONFIRMED. THE COMPLAINT WAS INTERPRETED AS "SEAL DID NOT DEPLOY PROPERLY". A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III PROXIMAL SEAL FELL APART UPON PULLING IT OUT OF THE AORTA. IT DID NOT OPEN UP CORRECTLY. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PT EFFECTS. THE PRODUCT IS RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTRING III PROXIMAL SEAL SYSTEM, 4.3 MM PROXIMAL SEAL DXC MAQUET CARDIOVASCULAR, LLC HSK-3043 25015050

Patients

Seq Age Sex Outcome Treatment
1 NA