HEARTSTRING III PROXIMAL SEAL SYSTEM, 4.3 MM
Report
- Report Number
- 2242352-2010-03519
- Event Type
- Malfunction
- Date Received
- December 3, 2010
- Date of Event
- November 4, 2010
- Report Date
- November 5, 2010
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- DXC
- PMA / PMN Number
- K080169
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION SUMMARY: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED INSIDE THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE INSIDE THE BODY OF THE LOADING DEVICE. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE NOT DEPRESSED ON THE DELIVERY DEVICE. THERE WAS NO EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL DID NOT LOAD PROPERLY" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).
THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III PROXIMAL SEAL DID NOT LOAD PROPERLY. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PT EFFECTS. THE PRODUCT IS RETURNING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSTRING III PROXIMAL SEAL SYSTEM, 4.3 MM | PROXIMAL SEAL | DXC | MAQUET CARDIOVASCULAR, LLC | HSK-3043 | 25018576 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |