FDA Adverse Event Malfunction Summary report: N

TERUMO TCM II COOLING AND HEATING SYSTEM

MDR report key: 1935821 · Received December 3, 2010

Report

Report Number
1828100-2010-02363
Event Type
Malfunction
Date Received
December 3, 2010
Date of Event
November 4, 2010
Report Date
December 3, 2010
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
K883603
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TC
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

DURING PREPARATION FOR USE OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE USER REPORTED NO FLOW IN THE HEATER COOLER. THE USER REPORTED THAT THE IMPELLER ASSEMBLY AND CERAMIC SPINDLE OF THE MAIN MOTOR PUMP WERE BROKEN. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED FOR THE PROCEDURE. THE USER REPORTED THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WERE NO ADVERSE CONSEQUENCES TO A PATIENT AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TERUMO TCM II COOLING AND HEATING SYSTEM HEATING COOLING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 4416

Patients

Seq Age Sex Outcome Treatment
1