TRUE METRIX

MDR report key: 19358088 · Received May 21, 2024

Report

Report Number
1000113657-2024-00227
Event Type
Malfunction
Date Received
May 21, 2024
Date of Event
April 28, 2024
Report Date
May 21, 2024
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER WAS RETURNED FOR EVALUATION. PRODUCT TESTING WAS PERFORMED AND NO DEFECT FOUND ON RETURNED METER. TEST STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-020: USER'S TEST STRIP HAD POOR STORAGE. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 13-MAY-2024 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED REPLACEMENT PRODUCTS RESOLVED INITIAL CONCERN.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 173, 334, 181 AND 165 MG/DL. CUSTOMER ALSO REPORTED THAT HE HAD OBTAINED A RESULT OF 354 MG/DL (UNDISCLOSED IF FASTING OR NON-FASTING) AND HE HAD ADMINISTERED 5 UNITS OF INSULIN. CUSTOMER STATED WHILE OUT WITH HIS FAMILY HE HAD ALMOST PASSED OUT. CUSTOMER STATED HE HAD EATEN AND FELT BETTER. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS AND REPORTED SYMPTOMS. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT IS BELOW 150 MG/DL. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION (BATHROOM). THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 02/24/2025 AND TEST STRIPS WERE OPEN 2 DAYS PRIOR TO THE CALL. THE CUSTOMER DID NOT HAVE ANOTHER VIAL OF TEST STRIPS THAT HAD BEEN STORED AND HANDLED CORRECTLY. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE/TIME NOT SET): RESULT 1: 173 MG/DL DATE: 5/28/24 TIME: 6:27 PM FASTING RESULT 2: 334 MG/DL DATE: 5/28/24 TIME: 3:10 PM FASTING RESULT 3: 106 MG/DL DATE: 5/28/24 TIME: 2;16 AM FASTING RESULT 4: 181 MG/DL DATE: 5/28/24 TIME: 12:19 AM FASTING RESULT 5: 165 MG/DL DATE: 5/27/24 TIME: 1:00 PM FASTING AM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1254081 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TMX WGN 100CT12/CS MG/DL #383569 ZB5341S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown