HBO VENT, MODEL HV-500A
Report
- Report Number
- 2020676-2010-00033
- Event Type
- Malfunction
- Date Received
- December 3, 2010
- Date of Event
- November 5, 2010
- Report Date
- December 3, 2010
- Manufacturer
- SECHRIST INDUSTRIES, INC.
- Product Code
- CBK
- PMA / PMN Number
- K832508
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
DEVICE RECEIVED AND IS CURRENTLY UNDER EVALUATION. RESULTS TO BE PROVIDED WHEN COMPLETED.
DURING THE DECOMPRESSION STATE IN THE CHAMBER THE EXHALATION PHASE ON THE HBO VENTILATOR STUCK CAUSING IT TO STAY CLOSED. THIS IN TURN CAUSED THE VENT TO NOT CYCLE OVER THE EXHALATION RESULTING IN PROLONGED ELEVATED AIRWAY PRESSURES. PERFUSIONIST REPORTED HE IS NOT SURE WHICH PATIENT USED WITH WHICH VENT. TREATMENT DATE: (B)(6) 2010. TREATED FOR CARBON MONOXIDE POISONING. NO CARDIAC EMBARRASSMENT (SATURATION LEVELS DID NOT DROP OFF). THE PROBLEM WAS CAUGHT QUICKLY AND VENTILATOR WAS CUT-OFF AND THEN WAS MANUALLY VENTILATED. THE VENTILATOR WAS IMMEDIATELY PULLED OUT OF SERVICE.
DEVICE RECEIVED AND IS CURRENTLY UNDER EVALUATION. RESULTS TO BE PROVIDED WHEN COMPLETED.
DURING THE DECOMPRESSION STATE IN THE CHAMBER THE EXHALATION PHASE ON THE HBO VENTILATOR STUCK CAUSING IT TO STAY CLOSED. THIS IN TURN CAUSED THE VENT TO NOT CYCLE OVER THE EXHALATION RESULTING IN PROLONGED ELEVATED AIRWAY PRESSURES. PERFUSIONIST REPORTED HE IS NOT SURE WHICH PATIENT USED WITH WHICH VENT. TREATMENT DATE: (B)(6) 2010. TREATED FOR CARBON MONOXIDE POISONING. THE PATIENT DID DEVELOP ARRYTHMIA CARDIAC AND HAD TO BE MANUALLY VENTILATED WHEN SHE WAS BROUGHT OUT OF THE CHAMBER SO THAT HER HEART RATE COULD BE BROUGHT BACK UP. HER HEART RATE WENT DOWN TO 50.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HBO VENT, MODEL HV-500A | NONE | CBK | SECHRIST INDUSTRIES, INC. | 500A VENT |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR | Required Intervention | ||
| 2 | 74 YR |