FDA Adverse Event
Injury
Summary report: N
PITT EASY ABUTMENT
MDR report key: 1935805
·
Received December 22, 2010
Report
- Report Number
- 9611820-2010-00023
- Event Type
- Injury
- Date Received
- December 22, 2010
- Report Date
- November 22, 2010
- Manufacturer
- SYBRON IMPLANT SOLUTIONS GMBH
- Product Code
- NHA
- PMA / PMN Number
- K053242
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
THE ABUTMENT WAS RETURNED TO (B)(4) AND EVALUATED. A VISUAL INSPECTION INDICATED THAT THE ABUTMENT MEETS PRODUCT SPECIFICATIONS AND THAT THE HEX FRACTURE WAS NOT THE RESULT OF A DESIGN FLAW OR A MANUFACTURING ERROR, BUT TO OVERLOADING AFTER SCREW LOOSENING. THIS IS NOT A PRODUCT FAILURE.
Description of Event or Problem · 1
ON (B)(6), 2010 A DOCTOR REPORTED TO (B)(4) THAT A PITT EASY ABUTMENT FRACTURED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PITT EASY ABUTMENT | ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS | NHA | SYBRON IMPLANT SOLUTIONS GMBH | 949201 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other| R |