FDA Adverse Event Injury Summary report: N

PITT EASY ABUTMENT

MDR report key: 1935805 · Received December 22, 2010

Report

Report Number
9611820-2010-00023
Event Type
Injury
Date Received
December 22, 2010
Report Date
November 22, 2010
Manufacturer
SYBRON IMPLANT SOLUTIONS GMBH
Product Code
NHA
PMA / PMN Number
K053242
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE ABUTMENT WAS RETURNED TO (B)(4) AND EVALUATED. A VISUAL INSPECTION INDICATED THAT THE ABUTMENT MEETS PRODUCT SPECIFICATIONS AND THAT THE HEX FRACTURE WAS NOT THE RESULT OF A DESIGN FLAW OR A MANUFACTURING ERROR, BUT TO OVERLOADING AFTER SCREW LOOSENING. THIS IS NOT A PRODUCT FAILURE.

Description of Event or Problem · 1

ON (B)(6), 2010 A DOCTOR REPORTED TO (B)(4) THAT A PITT EASY ABUTMENT FRACTURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PITT EASY ABUTMENT ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS NHA SYBRON IMPLANT SOLUTIONS GMBH 949201

Patients

Seq Age Sex Outcome Treatment
1 Other| R