SYNCHRON® LX20 ANALYZER (WITHOUT CAP PIERCER)
MDR report key: 1935629
·
Received December 22, 2010
Report
- Report Number
- 2050012-2010-01716
- Event Type
- Malfunction
- Date Received
- December 22, 2010
- Date of Event
- December 1, 2010
- Report Date
- December 1, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K011213
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) WENT ON-SITE AND FOUND A DEFECTIVE FLOAT SWITCH ON THE WASH SOLUTION CANISTER WHICH CAUSED THE CANISTER TO OVERFLOW AND CAUSED THE SOLUTION TO GET IN THE REGULATORS AND LINES. FSE REPLACED THE FLOAT SWITCH AND THE SYSTEM WAS RESUMED.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REPORTING FLUID LEAK FROM HYDRO LINE. THE REAGENT CONNECTED TO THIS LINE WAS WASH SOLUTION. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON® LX20 ANALYZER (WITHOUT CAP PIERCER) | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | NA | NA |