FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK 20 DEFIBRILLATOR/MONITOR
MDR report key: 1935483
·
Received December 14, 2010
Report
- Report Number
- 3015876-2010-01363
- Event Type
- Malfunction
- Date Received
- December 14, 2010
- Date of Event
- November 15, 2010
- Report Date
- November 15, 2010
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- LDD
- PMA / PMN Number
- K012274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED EVENT CODES LOGGED IN THE MEMORY. PHYSIO REPLACED THE THERAPY PCB ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE ILLUMINATED THE SERVICE INDICATOR AND LOGGED SEVERAL EVENT CODES THAT INDICATE A CRITICAL FAILURE IN THE MEMORY. THERE WAS NO PT USE ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 20 DEFIBRILLATOR/MONITOR | LDD | PHYSIO-CONTROL, INC. | 20 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |