ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2010-02880
- Event Type
- Malfunction
- Date Received
- December 14, 2010
- Report Date
- November 17, 2010
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARS TO BE INTACT, NO LIFTING OR PEELING. THE "UP/DOWN/OK" KEYPAD BUTTONS ARE INTERMITTENTLY RESPONDING TO USER INPUTS DURING TESTING. THE "CONTRAST" KEYPAD BUTTON REQUIRES EXCESSIVE FORCE TO ACTIVATE DURING TESTING. THE KEYPAD WAS REMOVED FOR FURTHER INVESTIGATION. DURING A VISUAL INSPECTION, THE "DOWN/CONTRAST" KEY CONTACTS ARE MISALIGNED, THE "CONTRAST" KEY CONTACT IS INVERTED, AND THE "UP/DOWN/OK" KEY CONTACTS BECOME INVERTED WHEN PRESSED AND DO NOT ALWAYS SPRING BACK. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
THE LAY-USER/PT ALLEGED THAT THE BUTTONS ON THE KEYPAD DO NOT RESPOND. THERE WAS PRODUCT MISUSE. THE KEYPAD DID NOT APPEAR TO BE PEELED OR TORN. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |