FDA Adverse Event Malfunction Summary report: N

FIXED DURAGUARD, 16MM

MDR report key: 1935467 · Received December 13, 2010

Report

Report Number
1811755-2010-02006
Event Type
Malfunction
Date Received
December 13, 2010
Date of Event
November 3, 2010
Report Date
November 4, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DRILL ATTACHMENT WAS RETURNED TO THE MFR FOR AN UNRELATED ISSUE AND UPON EVAL, AN OVERHEATING CONDITION WAS DISCOVERED. INTERNAL COMPONENTS WERE EVALUATED AND EXTENSIVE DEBRIS AND CORROSION WAS DISCOVERED WITHIN THE UPPER BEARINGS. THE PRESENCE OF DEBRIS/CORROSION CAN LEAD TO INCREASED FRICTION BETWEEN ROTATING COMPONENTS, RESULTING IN HEAT BEING GENERATED. IMPROPER OR INADEQUATE CLEANING/STERILIZATION TECHNIQUES COULD CONTRIBUTE TO THE BUILD-UP OF DEBRIS/CORROSION WITHIN THIS DEVICE. THE DRILL ATTACHMENT COULD NOT BE REPAIRED AND WAS DISCARDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MFR, THE DRILL ATTACHMENT HEATED UP. THIS EVENT DID NOT OCCUR IN A SURGICAL ENVIRONMENT, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FIXED DURAGUARD, 16MM DRILLS, BURS, TREPHINES & ACCESSORIES (SIMPLE, POW HBE STRYKER INSTRUMENTS KALAMAZOO 10071

Patients

Seq Age Sex Outcome Treatment
1 UNK