FIXED DURAGUARD, 16MM
Report
- Report Number
- 1811755-2010-02006
- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Date of Event
- November 3, 2010
- Report Date
- November 4, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HBE
- PMA / PMN Number
- K032303
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- UNKNOWN
Narratives
THE DRILL ATTACHMENT WAS RETURNED TO THE MFR FOR AN UNRELATED ISSUE AND UPON EVAL, AN OVERHEATING CONDITION WAS DISCOVERED. INTERNAL COMPONENTS WERE EVALUATED AND EXTENSIVE DEBRIS AND CORROSION WAS DISCOVERED WITHIN THE UPPER BEARINGS. THE PRESENCE OF DEBRIS/CORROSION CAN LEAD TO INCREASED FRICTION BETWEEN ROTATING COMPONENTS, RESULTING IN HEAT BEING GENERATED. IMPROPER OR INADEQUATE CLEANING/STERILIZATION TECHNIQUES COULD CONTRIBUTE TO THE BUILD-UP OF DEBRIS/CORROSION WITHIN THIS DEVICE. THE DRILL ATTACHMENT COULD NOT BE REPAIRED AND WAS DISCARDED.
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MFR, THE DRILL ATTACHMENT HEATED UP. THIS EVENT DID NOT OCCUR IN A SURGICAL ENVIRONMENT, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FIXED DURAGUARD, 16MM | DRILLS, BURS, TREPHINES & ACCESSORIES (SIMPLE, POW | HBE | STRYKER INSTRUMENTS KALAMAZOO | 10071 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |