FDA Adverse Event Malfunction Summary report: N

INTRAMATIC LUX3 CONTRA ANGLE 25LHA

MDR report key: 1935412 · Received November 11, 2010

Report

Report Number
3003637274-2010-00043
Event Type
Malfunction
Date Received
November 11, 2010
Date of Event
October 4, 2010
Report Date
October 11, 2010
Manufacturer
KAVO DENTAL GMBH
Product Code
EFB
PMA / PMN Number
K073478
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE HANDPIECE GETS WARM. THE PT WAS NOT BURNED BY THE HANDPIECE. DURING PRODUCT EVALUATION WAS FOUND, (B)(6) DOES SLIP (RUNS OUT OF SPEC), THE WATER SPRAY HOLES WAS CLOGGED. THE BEARINGS WHERE GRITTY AND THE HEAD OF THE HANDPIECE GET'S WARM NOT HOT.

Description of Event or Problem · 1

A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE HANDPIECE GET'S WARM. THE PT WAS NOT BURNED BY THE HANDPIECE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTRAMATIC LUX3 CONTRA ANGLE 25LHA DENTAL HANDPIECE EFB KAVO DENTAL GMBH 25LHA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention