FDA Adverse Event
Malfunction
Summary report: N
INTRAMATIC LUX3 CONTRA ANGLE 25LHA
MDR report key: 1935412
·
Received November 11, 2010
Report
- Report Number
- 3003637274-2010-00043
- Event Type
- Malfunction
- Date Received
- November 11, 2010
- Date of Event
- October 4, 2010
- Report Date
- October 11, 2010
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE HANDPIECE GETS WARM. THE PT WAS NOT BURNED BY THE HANDPIECE. DURING PRODUCT EVALUATION WAS FOUND, (B)(6) DOES SLIP (RUNS OUT OF SPEC), THE WATER SPRAY HOLES WAS CLOGGED. THE BEARINGS WHERE GRITTY AND THE HEAD OF THE HANDPIECE GET'S WARM NOT HOT.
Description of Event or Problem · 1
A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE HANDPIECE GET'S WARM. THE PT WAS NOT BURNED BY THE HANDPIECE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTRAMATIC LUX3 CONTRA ANGLE 25LHA | DENTAL HANDPIECE | EFB | KAVO DENTAL GMBH | 25LHA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |