ATHLETIS PTA BALLOON DILATATION CATHETER
Report
- Report Number
- 2124215-2024-30174
- Event Type
- Malfunction
- Date Received
- May 20, 2024
- Date of Event
- April 30, 2024
- Report Date
- June 20, 2024
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- LIT
- UDI-DI
- 08714729974260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
D2B - PRO CODE (PRODUCT CODE): LIT, DQY.
D2B PRO CODE (PRODUCT CODE): LIT, DQY.
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 12.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, BEFORE THE BALLOON REACHED MEDIUM PRESSURE, IT RUPTURED. THE DEVICE WAS TAKEN OUT AND THE PATIENT WAS RE-TREATED FOR ANOTHER OF THE SAME DEVICE TO COMPLETE THE PROCEDURE. THERE WERE NO ABNORMALITIES OR PATIENT COMPLICATIONS REPORTED.
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 12.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, BEFORE THE BALLOON REACHED MEDIUM PRESSURE, IT RUPTURED. THE DEVICE WAS TAKEN OUT AND THE PATIENT WAS RE-TREATED FOR ANOTHER OF THE SAME DEVICE TO COMPLETE THE PROCEDURE. THERE WERE NO ABNORMALITIES OR PATIENT COMPLICATIONS REPORTED. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS 70% STENOSED AND WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED CEPHALIC VEIN. THE BALLOON RUPTURED AT 12 ATMOSPHERES FOR 30 SECONDS. THE PATIENT'S CURRENT STATUS IS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1788466 | ATHLETIS PTA BALLOON DILATATION CATHETER | CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL | LIT | BOSTON SCIENTIFIC CORPORATION | 39347-120470 | 0029389743 | 08714729974260 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Male |