FDA Adverse Event Malfunction Summary report: N

ATHLETIS PTA BALLOON DILATATION CATHETER

MDR report key: 19353979 · Received May 20, 2024

Report

Report Number
2124215-2024-30174
Event Type
Malfunction
Date Received
May 20, 2024
Date of Event
April 30, 2024
Report Date
June 20, 2024
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
LIT
UDI-DI
08714729974260
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D2B - PRO CODE (PRODUCT CODE): LIT, DQY.

Additional Manufacturer Narrative · 0

D2B PRO CODE (PRODUCT CODE): LIT, DQY.

Description of Event or Problem · 0

IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 12.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, BEFORE THE BALLOON REACHED MEDIUM PRESSURE, IT RUPTURED. THE DEVICE WAS TAKEN OUT AND THE PATIENT WAS RE-TREATED FOR ANOTHER OF THE SAME DEVICE TO COMPLETE THE PROCEDURE. THERE WERE NO ABNORMALITIES OR PATIENT COMPLICATIONS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 12.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, BEFORE THE BALLOON REACHED MEDIUM PRESSURE, IT RUPTURED. THE DEVICE WAS TAKEN OUT AND THE PATIENT WAS RE-TREATED FOR ANOTHER OF THE SAME DEVICE TO COMPLETE THE PROCEDURE. THERE WERE NO ABNORMALITIES OR PATIENT COMPLICATIONS REPORTED. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS 70% STENOSED AND WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED CEPHALIC VEIN. THE BALLOON RUPTURED AT 12 ATMOSPHERES FOR 30 SECONDS. THE PATIENT'S CURRENT STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1788466 ATHLETIS PTA BALLOON DILATATION CATHETER CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC CORPORATION 39347-120470 0029389743 08714729974260

Patients

Seq Age Sex Outcome Treatment
1 69 YR Male