LIFEVEST WCD 3100 SYSTEM
Report
- Report Number
- 3002158293-2010-01302
- Event Type
- Malfunction
- Date Received
- December 7, 2010
- Date of Event
- July 22, 2010
- Report Date
- December 4, 2010
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY PACK SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE CAUSE OF THE DISCHARGED BATTERY BACK WAS DEFECTIVE CELLS WITHIN THE BATTERY PACK. ONE OF THE CELLS HAD RESIDUE/CORROSION ON THEM. THE SOURCE OF THE RESIDUE/CORROSION HAS NOT BEEN POSITIVELY IDENTIFIED BUT IS LIKELY ELECTROLYTE FROM A LEAKING CELL. THE ROOT CAUSE OF THE LEAKING CELL WAS NOT POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PT RECEIVED A REPLACEMENT BATTERY PACK.
THE TERRITORY MANAGER ASSISTING A (B)(6) OLD MALE PT CALLED IN TO ZOLL LIFECOR CUSTOMER SUPPORT TO REPORT ONE OF THE PATIENT'S BATTERIES IS DISPLAYING RED FAULT LIGHTS WHEN PLACED ON THE CHARGER AND IS UNABLE TO POWER ON THE MONITOR. SUPPORT ISSUED THE PT A REPLACEMENT BATTERY PACK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3100 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |