FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3100 SYSTEM
MDR report key: 1935188
·
Received December 7, 2010
Report
- Report Number
- 3002158293-2010-01301
- Event Type
- Malfunction
- Date Received
- December 7, 2010
- Date of Event
- June 22, 2010
- Report Date
- December 4, 2010
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY PACK (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE CAUSE OF THE DISCHARGED BATTERY PACK WAS DEFECTIVE CELLS WITHIN THE BATTERY PACK. ONE OF THE CELLS HAD RESIDUE/CORROSION ON THEM. THE SOURCE OF THE RESIDUE/CORROSION HAS NOT BEEN POSITIVELY IDENTIFIED BUT IS LIKELY ELECTROLYTE FROM A LEAKING CELL. THE ROOT CAUSE OF THE LEAKING CELL WAS NOT POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PT REC'D A REPLACEMENT BATTERY PACK.
Description of Event or Problem · 1
A (B)(6) MALE PT'S WIFE CALLED IN TO ZOLL LIFECOR CUSTOMER SUPPORT TO REPORT THAT THE PT'S CHARGER IS FLASHING A RED LIGHT WHEN THE BATTERY WAS PLACED IN IT. SUPPORT ISSUED THE PT A REPLACEMENT BATTERY PACK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3100 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR |