REPAIR CORE RECIP SAW
MDR report key: 1935150
·
Received December 6, 2010
Report
- Report Number
- 1811755-2010-01917
- Event Type
- Malfunction
- Date Received
- December 6, 2010
- Date of Event
- November 8, 2010
- Report Date
- November 8, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE WAS RECEIVED AT THE MANUFACTURER FOR EVALUATION, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS CONFIRMED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS PROBLEMS WITH A WORN BUTTON AND SLIDE LOCK, BOTH OF WHICH WERE REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE OVERHEATED DURING ROUTINE PREVENTATIVE MAINTENANCE TESTING BY THE MANUFACTURER. THIS DID NOT OCCUR DURING ANY SURGICAL OR MEDICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO ADVERSE CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | REPAIR CORE RECIP SAW | SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTACHMENT | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |