FDA Adverse Event Malfunction Summary report: N

REPAIR CORE RECIP SAW

MDR report key: 1935150 · Received December 6, 2010

Report

Report Number
1811755-2010-01917
Event Type
Malfunction
Date Received
December 6, 2010
Date of Event
November 8, 2010
Report Date
November 8, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HANDPIECE WAS RECEIVED AT THE MANUFACTURER FOR EVALUATION, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS CONFIRMED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS PROBLEMS WITH A WORN BUTTON AND SLIDE LOCK, BOTH OF WHICH WERE REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HANDPIECE OVERHEATED DURING ROUTINE PREVENTATIVE MAINTENANCE TESTING BY THE MANUFACTURER. THIS DID NOT OCCUR DURING ANY SURGICAL OR MEDICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO ADVERSE CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REPAIR CORE RECIP SAW SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTACHMENT DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK